Key facts
Start here

Agilent 850-DS Dissolution Sampling Station Method Setup Quick Start Guide
This quick start guide describes how the Agilent 850-DS Dissolution Sampling Station integrates with existing workflows to give unattended sampling and cleaning, with the aim of raising analysis efficiency and removing variability. The station supports several dissolution methods and can be used with USP Apparatus 1, 2, 3, 5, 6 and 7 dissolution testers, and results can be collected in traditional test tubes, HPLC vials or 96-well plates.
The document is built around a key parameter table that lists fill volume, purge volume, waste volume, estimated minimum transfer time, aspiration dwell time, pump speed and fill loss volume, together with the importance of each parameter, setup guidance and minimum, typical and maximum values. It also points to the 850-DS cleaning procedure, volume calibration procedure and user manual, and gives the Agilent dissolution hotline e-mail address and community link for further method setup help. This manual is written in Chinese.
Agilent 850-DS specifications
| Supported dissolution apparatus | USP Apparatus 1, 2, 3, 5, 6 and 7 |
|---|---|
| Sample containers | Traditional test tubes, HPLC vials or 96-well plates |
| Fill volume range | 4 mL min / 10 mL typical / 25 mL max |
| Purge volume range | 6 mL min / 12 mL typical / 25 mL max |
| Waste volume range | 0.2 mL min / 0.5 mL typical / 10 mL max |
| Aspiration dwell time range | 1 s min / 1 s typical / 5 s max |
| Minimum aspiration dwell time with filter module | 3 s |
| Pump speed options | 6, 8, 10 or 12 mL/min |
| Fill loss volume | 4 mL (708-DS/850-DS); 7 mL (708-DS/850-DS/Cary 60) |
Source: Agilent 850-DS Dissolution Sampling Station Method Setup Quick Start Guide
All Agilent 850-DS documents by type
The shortest path to a working setup, usually only a few pages.
Questions answered in these manuals
Which USP dissolution apparatus can the 850-DS be used with?
It can be used with USP Apparatus 1, 2, 3, 5, 6 and 7 dissolution testers.
From: Agilent 850-DS Dissolution Sampling Station Method Setup Quick Start Guide
Does the 850-DS clean itself after a dissolution test?
Yes. The 850-DS can perform self-cleaning after each dissolution test.
From: Agilent 850-DS Dissolution Sampling Station Method Setup Quick Start Guide
Which USP dissolution apparatuses does the Agilent 850-DS work with?
It works with USP dissolution apparatus 1, 2, 3, 5, 6, and 7.
From: Agilent 850-DS Dissolution Sampling Station Practical Solutions Newsletter
What sample containers can the 850-DS use?
It can use conventional test tubes, HPLC vials, or 96-well plates.
From: Agilent 850-DS Dissolution Sampling Station Practical Solutions Newsletter