Agilent 4150/4200 TapeStation High Sensitivity RNA Quick Start Guide
This simple manual (ver. 3.0, April 2021) explains how to run High Sensitivity RNA assays on the Agilent 4150/4200 TapeStation. It lists required reagents and consumables, including ScreenTape, sample buffer, ladder, loading tips, PCR tube strips, plates, and seals, and states that software Revision 4.1 or later is required.
Setup steps cover loading ScreenTape, selecting 8-tube strips or a 96-well plate, and using the Controller software to choose sample wells. The document also gives sample concentration ranges, notes that only specified consumables should be used, and refers to a separate troubleshooting guide for needle replacement and maintenance. This manual is written in Japanese.
Agilent 4200 TapeStation specifications
| Quantitation range | 500-10000 pg/μL |
|---|---|
| RINe recommended concentration range | 1000-25000 pg/μL |
| Operating temperature | 14-30°C |
| Optimal temperature | 20°C |
Source: Agilent 4150/4200 TapeStation High Sensitivity RNA Quick Start Guide
All Agilent 4200 TapeStation documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
The shortest path to a working setup, usually only a few pages.
Questions answered in these manuals
Where can I find the user manuals for the Agilent 4200 TapeStation?
User manuals are provided on the included USB flash drive, in printed form, or can be found on the Agilent website by searching for the product number.
From: Agilent 4200 TapeStation System Software & User Information
What should be done before applying power to the instrument?
You must verify that all safety precautions from the individual instrument's user manual are taken and that the power distribution's frequency and voltage range match the instrument's power specification.
From: Agilent 4200 TapeStation System Software & User Information
What reagents are required for High Sensitivity RNA analysis?
ScreenTape and the reagent kit are required.
From: Agilent 4150/4200 TapeStation High Sensitivity RNA Quick Start Guide
Can loading tips for the 2200 TapeStation be used?
No, 2200 TapeStation loading tips cannot be used.
From: Agilent 4150/4200 TapeStation High Sensitivity RNA Quick Start Guide