Agilent 4150/4200 TapeStation High Sensitivity RNA quick start guide covering required reagents, consumables, setup, software requirements, and sample…
This simple manual (ver. 3.0, April 2021) explains how to run High Sensitivity RNA assays on the Agilent 4150/4200 TapeStation. It lists required reagents and consumables, including ScreenTape, sample buffer, ladder, loading tips, PCR tube strips, plates, and seals, and states that software Revision 4.1 or later is required.
Setup steps cover loading ScreenTape, selecting 8-tube strips or a 96-well plate, and using the Controller software to choose sample wells. The document also gives sample concentration ranges, notes that only specified consumables should be used, and refers to a separate troubleshooting guide for needle replacement and maintenance. This manual is written in Japanese.
| Quantitation range | 500-10000 pg/μL |
|---|---|
| RINe recommended concentration range | 1000-25000 pg/μL |
| Operating temperature | 14-30°C |
| Optimal temperature | 20°C |
ScreenTape and the reagent kit are required.
No, 2200 TapeStation loading tips cannot be used.
No, the ladder is not necessary if only the degradation score (RINe) is measured.
The operating environment temperature is 14-30°C.