Agilent 4150/4200 TapeStation High Sensitivity RNA Quick Start Guide

Agilent 4150/4200 TapeStation High Sensitivity RNA quick start guide covering required reagents, consumables, setup, software requirements, and sample…


Agilent 4150/4200 TapeStation High Sensitivity RNA Quick Start Guide - cover page
Brand
Agilent
Category
Analyzer
Document type
Quick Start Guide
Language
Japanese
Pages
14
File format
PDF
File size
2 MB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
9A90F2221E1C00BBE3BB073E47B63FED

About this manual

This simple manual (ver. 3.0, April 2021) explains how to run High Sensitivity RNA assays on the Agilent 4150/4200 TapeStation. It lists required reagents and consumables, including ScreenTape, sample buffer, ladder, loading tips, PCR tube strips, plates, and seals, and states that software Revision 4.1 or later is required.

Setup steps cover loading ScreenTape, selecting 8-tube strips or a 96-well plate, and using the Controller software to choose sample wells. The document also gives sample concentration ranges, notes that only specified consumables should be used, and refers to a separate troubleshooting guide for needle replacement and maintenance. This manual is written in Japanese.

What's inside

  • Required reagents and consumables
  • Before starting preparation
  • Device setup
  • ScreenTape setup
  • Sample preparation notes
  • Controller software settings
  • Troubleshooting guide
  • Maintenance

Specifications

Quantitation range500-10000 pg/μL
RINe recommended concentration range1000-25000 pg/μL
Operating temperature14-30°C
Optimal temperature20°C

Frequently asked questions

What reagents are required for High Sensitivity RNA analysis?

ScreenTape and the reagent kit are required.

Can loading tips for the 2200 TapeStation be used?

No, 2200 TapeStation loading tips cannot be used.

Is the RNA ladder required when only measuring RINe?

No, the ladder is not necessary if only the degradation score (RINe) is measured.

What is the operating temperature range?

The operating environment temperature is 14-30°C.

Download PDF (2 MB)