Agilent 1000/2000 Manuals

ManualLib has 2 documents for the Agilent 1000/2000: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Brand
Agilent
Model
1000/2000
Industry
Industrial
Documents
2
Languages
EN
Last added
2026-11-21

Start here

Agilent Flow Tracker Series 1000/2000 Reference Manual

Agilent Flow Tracker Series 1000/2000 Reference Manual

User Manual 36 pages · 989 KB · EN

The Agilent Flow Tracker Series 1000 and 2000 are portable instruments designed to provide simple, reliable, and portable flow information for gas chromatography (GC). The Flow Tracker 1000 operates in Flow mode and GC mode, while the Flow Tracker 2000 adds a Leak Detector mode and includes a built-in sampling pump and probe. Both models feature a multi-line LCD display for viewing parameters and an RS-232 output for real-time data streaming to a laptop computer.

Read online

Agilent 1000/2000 specifications

Operating temperature range10—50 °C
Operating humidity range0—90% RH (noncondensing)
Maximum recommended altitude15,000 ft
Maximum operating line pressure100 PSIG (690 kPa)
Power sourcesix (6) standard AA size batteries or optional AC/DC adapter

Source: Agilent Flow Tracker Series 1000/2000 Reference Manual

All Agilent 1000/2000 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the operating principle of the Flow Tracker?

The volumetric flow rate is determined by creating a pressure drop across a unique internal restriction known as a Laminar Flow Element and measuring differential pressure across it.

From: Agilent Flow Tracker Series 1000/2000 Reference Manual

What are the operating modes of the Flow Tracker 1000?

The Flow Tracker 1000 has two operating modes: Flow mode and Gas Chromatography (GC) mode.

From: Agilent Flow Tracker Series 1000/2000 Reference Manual

What are the operating modes of the Flow Tracker?

The 1000 Series has two operating modes (Flow Mode and GC Mode), while the 2000 Series adds a Leak Detector Mode.

From: Agilent Flow Tracker 1000/2000-Detailed guide

Can the Flow Tracker be used in hazardous locations?

No, this product is designed for indoor use only and should not be used in Hazardous Locations.

From: Agilent Flow Tracker 1000/2000-Detailed guide