Vaisala HMI41 Calibration Manual for HMD/W20/30 and HMP130 Series
The manual explains how to use the HMI41 with its calibration option to adjust analog measurement transducers of the HMD/W20/30 and HMP130 series. It provides details on the required calibration cables, the selection of calibration functions, and the step‑by‑step procedure for performing a calibration.
The HMI41 can operate in four modes, with function 4 dedicated to calibrating the mentioned analog transducers. The document also covers setup entry, display navigation, and the expected stabilization time for accurate readings. This manual is written in German.
All Vaisala HMDW2030 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How do I select the HMI41 calibrator function for HMDW2030/HMP130 transmitters?
Press the ON/OFF button, then press ENTER and MODE repeatedly until function number 4 appears, turn the HMI41 off and on again to activate the calibrator.
Which calibration cable is required for HMP130 series transmitters?
Use calibration cable part number 1916522 for the HMD/W 20/30 and HMP130 series analogue transmitters.