VAISALA HMDW6070/HMP140 Manual
This guide explains how to use the HMI41 as a field calibrator for Vaisala HMP140 analogue humidity transmitters. It covers the selection of calibrator functions, scale selection modes, and step‑by‑step calibration procedures.
The manual also provides the standard transmitter settings, default HMI41 memory values, and detailed instructions for voltage and current output selections, ensuring accurate on‑site calibration of the HMP140 series.
All Vaisala HMDW6070 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How do you select the HMI41 calibrator function for HMP140 transmitters?
Press the ON/OFF button, then within 1–2 seconds press both ENTER and MODE buttons until the calibrator function text appears on the display.
What are the standard transmitter settings for the HMP140?
The HMP140 standard settings are 0…100 %RH humidity range, –40…460 °C temperature range, 0…5 V voltage output, and 0…20 mA current output.