Agilent ZORBAX SB-C8 Plant Hormones Rapid Gradient Elution Application Note

The document describes the features and characteristics of some plant growth regulators, including the types and mechanisms of plant hormones.


Agilent ZORBAX SB-C8 Plant Hormones Rapid Gradient Elution Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
204 KB
Published
03 July, 2012
Updated
28 September, 2023
MD5 checksum
3A82EA519C21F7700C4C664E81E7348C

About this manual

This document serves as an application brief for the rapid analysis of plant hormones using High-Performance Liquid Chromatography (HPLC). It details the separation of twelve specific plant hormone compounds, highlighting the utility of gradient elution to reduce runtime and increase throughput for research into plant metabolism and disease mechanisms.

The recommended method utilizes an Agilent ZORBAX SB-C8 column (4.6 x 75 mm) with a Hewlett Packard 1050 LC system. The separation is achieved using a gradient of water with 0.1% TFA and acetonitrile with 0.1% TFA, with the organic modifier increasing from 22% to 50% over 1.5 minutes. The analysis is performed at 60°C with a flow rate of 3.0 mL/min and detection at 254 nm.

What's inside

  • Application Brief
  • Detector Response
  • Retention Time
  • Conditions
  • Highlights
  • Plant Hormones

Specifications

ColumnZORBAX SB-C8, 4.6 x 75 mm
Column Part Number66953-906
Mobile Phase AWater with 0.1% TFA
Mobile Phase BAcetonitrile with 0.1% TFA
Gradient22% B to 50% B in 1.5 min
Wavelength254 nm
Flow Rate3.0 mL / min

Frequently asked questions

What compounds are separated in this application?

The method separates twelve plant hormones including Kinetin, Indole Acetic Acid, and various chlorophenoxy acids.

What is the stated length and diameter of the ZORBAX SB-C8 column?

The document specifies a column size of 4.6 x 75 mm.

What is the recommended operating temperature for this separation?

The procedure requires using a temperature of 60°C.

What is the specified flow rate for the mobile phase?

The application note states a flow rate of 3.0 mL/min.

Download PDF (204 KB)