Agilent 1100 Series Analysis of Androgen Drugs by HPLC Application Note

This document introduces the HPLC analysis method for androgen drugs


Agilent 1100 Series Analysis of Androgen Drugs by HPLC Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
29 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
23BDD549D677A65242A7F6CEB748BD04

About this manual

The document presents a High-Performance Liquid Chromatography (HPLC) method for analyzing two androgen drugs: testosterone and testosterone acetate. The method utilizes a reversed-phase column with UV detection at variable wavelengths to separate the compounds effectively.

The analysis is performed using an Agilent 1100 Series system, which includes a quaternary pump, thermostatted autosampler, thermostatted column compartment, and a variable wavelength detector. Specific performance metrics such as Limit of Detection (LOD), retention time precision, and area precision are provided for both compounds.

What's inside

  • Analysis of Androgen Drugs
  • HPLC Performance
  • Equipment Requirements
  • Chromatogram Conditions
  • Column Specifications

Specifications

Column dimensions4.6 x 125 mm Hypersil ODS, 5 um
Mobile phase Awater
Flow rate1.0 ml/min
Column temperature40 °C
Injection volume5 ul

Frequently asked questions

What equipment is required for this analysis?

The method uses an Agilent 1100 Series system including a quaternary pump, thermostatted autosampler, and variable wavelength detector.

What is the recommended column for this method?

The recommended column is Hypersil ODS, 5 um, with dimensions of 4.6 x 125 mm.

What is the Limit of Detection for testosterone?

The LOD for testosterone is 0.25 mg/l based on a signal-to-noise ratio of 2.

Download PDF (29 KB)