Agilent 1100 Series HPLC Analysis of Antitumor Drugs

This document introduces the conditions for analysis of anti-tumor drugs using the Agilent 1100 series HPLC systems.


Agilent 1100 Series HPLC Analysis of Antitumor Drugs - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
29 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
522DA82AC1F5230D5C44C871DA8EFDF3

About this manual

This document details a High-Performance Liquid Chromatography (HPLC) method for the analysis of the antitumor agent paclitaxel (Taxol). The method utilizes a reversed-phase column and UV detection to separate and quantify the compound. Key parameters include a mobile phase gradient of water and acetonitrile, a flow rate of 1.0 ml/min, and a column temperature of 25 °C. The analysis is performed on an Agilent 1100 Series system equipped with a quaternary pump, thermostatted autosampler, and variable wavelength detector.

What's inside

  • Analysis of Antitumor Drugs
  • Chromatogram of Paclitaxel
  • HPLC Performance
  • Equipment
  • Columns

Specifications

Column4x 125 mm Hypersil ODS, 5 ym
Mobile phase Awater
Mobile phase Bacetonitrile
Flow rate1.0 ml/min
Column temperature25 °C
Stop time12 min
Injection volume5 ul
Precision of RT (RSD)0.07
Precision of Area (RSD)0.79
Correlation factor1.00000

Frequently asked questions

What equipment is required for this analysis?

The method uses an Agilent 1100 Series system with a quaternary pump, thermostatted autosampler, thermostatted column compartment, and variable wavelength detector.

What is the recommended column for this method?

The recommended column is Hypersil ODS, 5 um, 4x 125 mm (Agilent part number 7992618-564).

Is a cooled autosampler necessary?

Yes, it is necessary to use a cooled autosampler, such as the Agilent 1100 Series thermostatted autosampler, to avoid sample decomposition.