Agilent 1100 Series Analysis of Quercetin and Kaempferol in Ginkgo Extract Application Note

This document describes the results of the separation of quercetin and kaempferol in ginkgo extract and tablets using HPLC.


Agilent 1100 Series Analysis of Quercetin and Kaempferol in Ginkgo Extract Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
54 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
48256FCC7B7FC37D1C87052E87AE4E69

About this manual

The document presents a High-Performance Liquid Chromatography (HPLC) method developed on the Agilent 1100 Series system for separating and quantifying quercetin and kaempferol in Ginkgo biloba extracts and tablets. The analysis utilizes a reversed-phase column with UV detection at 370 nm.

What's inside

  • Analysis of Quercetin and Kaempferol
  • HPLC Conditions
  • Extraction Procedure
  • Equipment Requirements
  • Column Specifications
  • Spectral Comparison

Specifications

Column4x 125 mm Hypersil ODS, 5 um
Mobile phase A0.5 % H3PO4 in water
Mobile phase Bmethanol
Flow rate2.0 ml/min
UV detector wavelength370 nm
Column compartment temperature25°C
Injection volume10 ul

Frequently asked questions

Can this method be used on older or newer systems?

Since the method was specifically developed on the Agilent 1100 Series system you might not be able to reproduce this analysis on an older system or even on a new system with lower performance.

What is the recommended guard cartridge?

The recommended guard cartridge is Hypersil ODS, 5 um, 4x 4mm (Agilent part number 7992618-504).