Agilent 1100 Series Analysis of Glucocorticoids by HPLC Application Note

This document describes the method of HPLC analysis of glucocorticoids.


Agilent 1100 Series Analysis of Glucocorticoids by HPLC Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
33 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
CD93DA701523041135E14308D4147A32

About this manual

The document details a High-Performance Liquid Chromatography (HPLC) method for separating and analyzing various glucocorticoid steroid hormones, including beclomethasone, hydrocortisone, prednisolone, and betamethasone. The method utilizes a reversed-phase column with UV detection at 254 nm.

What's inside

  • Analysis of Glucocorticoids
  • HPLC Conditions
  • Equipment List
  • Performance Data
  • Column Specifications

Specifications

Column4x 125 mm Hypersil ODS, 5 um
Mobile phase Awater
Flow rate1.0 ml/min
UV detector wavelength254 nm
Column temperature25 °C

Frequently asked questions

What is the injection volume?

The recommended injection volume for this method is 5 ul.

What is the purpose of setting the autosampler temperature to 4°C?

The autosampler temperature was set to 4°C to avoid decomposition of the samples.

At what wavelength does the UV detector operate for this analysis?

The variable wavelength detector operates at a standard cell setting of 254 nm.