Agilent ZORBAX StableBond-C18 Rapid-Resolution High-Speed Separation of Parabens Application Note

This document describes a high-speed HPLC method for separating parabens. The method uses a ZORBAX StableBond SB-C18 column and 0.1% H3PO4:ACN as the mobile phase. The run conditions are 2 mL/min flow rate, 1 µL injection volume, and 60℃ column temperature. The method can separate parabens in less than one minute.


Agilent ZORBAX StableBond-C18 Rapid-Resolution High-Speed Separation of Parabens Application Note - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
282 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
BE40030790A5E2D299EA4C4E6C84E3D8

About this manual

This document details a high-speed analytical method for separating parabens, which are preservatives used in the cosmetic and food industries. The method achieves analysis times of less than one minute using a Rapid Resolution SB-C18 column.

The procedure utilizes an Agilent 1100 HPLC system with a ZORBAX StableBond-C18 Rapid-Resolution cartridge column (4.6 x 30 mm). Operating conditions include a mobile phase of 0.1% H3PO4 in ACN (50:50), a flow rate of 2 mL/min, and detection at 254 nm. Elevated temperatures (60°C) are used to further reduce run time.

What's inside

  • Application
  • Highlights
  • Operating Conditions
  • Chromatograms

Specifications

Flow Rate2 mL/min
Detection Wavelength254 nm
Analysis TimeLess than one minute

Frequently asked questions

What is the analysis time for separating parabens with this column?

The application note states that an analysis time of less than one minute is accomplished.

At what temperature should the bottom of the HPLC system be set to reduce run time?

The operating conditions specify setting the bottom temperature to 60°C.