Agilent J&W DB-35ms UI Ultra Inert GC Column Application Note

This document introduces a method for pesticide analysis using an Agilent J&W DB-35ms UI 20 m × 0.18 mm, 0.18 µm column, which can effectively separate organochlorine and organophosphorus pesticides with good separation and reproducibility.


Agilent J&W DB-35ms UI Ultra Inert GC Column Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
1.2 MB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
BD7887A266E81D067BF0EF8D44027C17

About this manual

This document details the analysis of challenging pesticide samples using the Agilent J&W DB-35ms Ultra Inert (UI) gas chromatography column. The note highlights how the column's inertness prevents peak tailing and degradation of active analytes like endrin and DDT, which are common issues in standard systems. It also emphasizes the column's midpolar phase for improved selectivity, allowing for the baseline resolution of coeluting peaks such as chlorpyrifos-methyl and methyl parathion.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Chromatographic Conditions
  • Flow Path Supplies
  • Solutions and Standards
  • Spiked Sample Preparation
  • Results and Discussion
  • Figure 1
  • Figure 2

Specifications

Column Dimensions20 m x 0.18 mm, 0.18 µm
Part Number121-3822UI
Carrier GasHelium
Flow Rate1.3 mL/min
Oven Temperature Program50 °C (1.3 min), 50 °C/min to 135 °C, 15 °C/min to 200 °C, 20 °C/min to 310 °C (2.5 min), 5 min at 310 °C

Frequently asked questions

What is the part number for the Agilent J&W DB-35ms UI column?

The part number for the Agilent J&W DB-35ms UI column is 121-3822UI.

How does the DB-35ms UI column improve analysis of active analytes?

The Ultra Inert series helps minimize column activity so that difficult and active analytes can be consistently analyzed at trace levels without degradation.

What advantage does the DB-35ms UI phase offer over a nonpolar phase?

The midpolar phase offers more selectivity to resolve difficult analytes, such as baseline-resolving chlorpyrifos-methyl and methyl parathion which coelute on 5% phenyl phases.