Varian CP-4900 DMD Micro-GC Application Note for Dichloroethane Analysis

This application note describes the analysis of very low levels (sub ppm) of 1,2-dichloroethane in which a special sample concentratior was used in conjunction with the Varian CP-4900 DMD Micro-GC with differential mobility detector.


Varian CP-4900 DMD Micro-GC Application Note for Dichloroethane Analysis - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
161 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
CP-4900
MD5 checksum
C6B37053CCFB26FB5FBA241F6CC5ABF4

About this manual

This document details a method for analyzing very low levels (sub-ppm) of 1,2-dichloroethane in air. It combines the Varian CP-4900 Micro-GC equipped with a Differential Mobility Detector (DMD) and an Enrichment and Desorption Unit (EDU) Sample Concentrator to achieve detection limits in the ppb range. The text explains that while the standard thermal conductivity detector (TCD) can detect low ppm levels, the DMD offers superior sensitivity and selectivity, further enhanced by sample enrichment.

What's inside

  • Introduction
  • Instrumentation
  • Conditions
  • Results and Discussion
  • Conclusion
  • Reference

Specifications

Desorption temperature200 °C
DMD Sensor temperature80 °C
DMD Flow rate450 mL/min
DMD Rf voltage1200 V
Oven temperature100 °C
Carrier gasHe

Frequently asked questions

What is the correlation coefficient for sample concentration linearity?

The linear behavior of sample concentration demonstrated a correlation coefficient of 0.9979 when increasing sampling time.

What is the repeatability of the system at low concentrations?

The results of 14 consecutive runs remain within a 10% RSD (Relative Standard Deviation).