Agilent 1200L LC/MS/MS Application Note for Cyclosporin A Analysis

The document evaluates the usefulness and performance of the Varian 1200L LC/MS/MS for the analysis of Cyclosporin A. The system includes automated on-line extraction and is controlled and fully automated through the Varian MS Workstation software. The Multi Reaction Monitoring (MRM) capability is used to demonstrate the sensitivity and robustness of the LC/MS/MS system.


Agilent 1200L LC/MS/MS Application Note for Cyclosporin A Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
394 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
84A60F56F370DB7989A2F6243438E37D

About this manual

This document evaluates the usefulness and performance of the Agilent 1200L LC/MS/MS system for analyzing Cyclosporin A. The study investigates the system's Multi Reaction Monitoring (MRM) capability, sensitivity, and robustness using automated on-line extraction and reporting via the Varian MS Workstation.

The analysis utilizes a Prostar 410 Autosampler and Prostar 210 Series LC pumps integrated with an API interface and a 1200L MS/MS triple quadrupole instrument. The study demonstrates the system's ability to produce a calibration curve with excellent linearity (r² = 0.999) and determine the Limit of Detection (LOD) and Limit of Quantitation (LOQ) for Cyclosporin A in bovine serum.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation/Procedure
  • HPLC Conditions
  • MS Conditions
  • Results and Discussion
  • Conclusion

Specifications

ColumnPursuit C18, 3 um, 100 x 2 mm
Column Temperature25°C
Injection Volume100 uL
Ionization ModeESI negative
Collision Gas2.0 mTorr Argon
API Drying Gas39 psi at 200°C
API Nebulizing Gas59 psi
SIM Width0.7 amu
Precursor Ion (Cyclosporin A)1200 m/z
Product Ion (Cyclosporin A)1089 m/z
Collision Energy (Cyclosporin A)38 V
Calibration Linearityr² = 0.999

Frequently asked questions

What compound was used as the test compound for this study?

The compound Cyclosporin A was employed as a test compound for this study.

What is the ionization mode used for the analysis?

The analysis uses ESI negative ionization mode.

What is the linearity of the calibration curve for Cyclosporin A?

Excellent linearity was found with an r² = 0.999.

What is the collision energy used for Cyclosporin A?

The collision energy used for Cyclosporin A is 38 V.