Agilent LC/MSD Ion Trap Application Note: Identification of Impurities in Drug Intermediates

This article describes the sensitive impurity detection of drug intermediates produced by Agilent Technologies, Inc. High purity standards are required for drug intermediates in the production of flurazepam, enalapril, and flunarizine. The Agilent LC/MSD Ion Trap was used to obtain LC/MS and MS/MS data for each intermediate, providing sufficient information to propose structures for the corresponding impurities.


Agilent LC/MSD Ion Trap Application Note: Identification of Impurities in Drug Intermediates - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
75 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
39BC0E9F573F9BE4C4F6BE4079DF1C61

About this manual

This document describes the application of the Agilent LC/MSD Ion Trap system for the sensitive detection and identification of impurities in fine chemicals and drug intermediates. The text highlights the importance of purity standards in drug production and the risks associated with impurities.

What's inside

  • Introduction
  • Experimental
  • Example 1
  • Example 2
  • Example 3
  • Conclusions
  • References

Specifications

System ConfigurationAgilent LC/MSD Ion Trap SL with 1100 binary pump and well plate autosampler
Sample Concentration5 to 10 mg/mL
Injection Volume1 uL
SmartFrag Range20% to 200% of fragmentation amplitude

Frequently asked questions

What capability allows for obtaining precursor ion and fragmentation in one run?

The Auto MS/MS capability with the application of the powerful SmartFrag allows for obtaining precursor ion and representative fragmentation in one LC/MS and MS/MS run.

Is MS/MS alone sufficient for positive identification of an unknown impurity?

No, MS/MS alone is insufficient for positive identification of an unknown; comparison with an authenticated reference is required.

What additional information provides confirmation of an impurity's identity?

Matching chromatographic retention time with an authenticated standard provides additional confirmation.