Varian 320-MS Triple Quadrupole LC/MS/MS Phospholipid Characterization Application Note

This application note demonstrates how phospholipids can be identified in a mixture by triple quadrupole mass spectrometry and suspected structures confirmed by generating MS/MS spectra and submitting it to the ACD Labs/MS Manager software.


Varian 320-MS Triple Quadrupole LC/MS/MS Phospholipid Characterization Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
245 KB
Published
12 June, 2012
Updated
28 September, 2023
Model
320-MS
MD5 checksum
1315B3DE1DE1F11F19D21B57E8A44113

About this manual

This document describes a method for identifying phospholipids in a mixture using triple quadrupole mass spectrometry. It details the use of the Varian 320-MS instrument equipped with an electrospray source and a built-in syringe infusion pump. The process involves preparing phospholipid standards (DPPE, POPG, and DPPG) and analyzing them via direct infusion.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • MS/MS Conditions
  • Discussion
  • Conclusion

Specifications

Collision Energy (DPPE)22V
Collision Energy (POPG)27 V
Collision Energy (DPPG)28 V
Infusion Flow Rate10 pL/min

Frequently asked questions

How are phospholipid samples prepared for this analysis?

Samples are prepared as pure standards suspended at a concentration of 1 ppm in a 50:50 methanol:water mixture for direct infusion.

What concentration of phospholipid standards was used for the direct infusion analysis?

The samples were prepared by suspending them at a concentration of 1 ppm in a 50:50 methanol:water mixture.

Download PDF (245 KB)