Varian 450-GC Analysis of Sulfur Components in LPGs Application Note

This document describes the analysis method for sulfur components in various LPGs. A highly selective pulsed flame photometric detector is used, and a dual-channel configuration is employed for the analysis. This method allows for the analysis of all components of interest in one run.


Varian 450-GC Analysis of Sulfur Components in LPGs Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
614 KB
Published
12 June, 2012
Updated
28 September, 2023
Model
450-GC
MD5 checksum
077B28BCE4CF016AF80A2B155E2058F4

About this manual

The document details a custom two-channel system developed for the analysis of sulfur-containing components such as H2S, COS, and mercaptanes. The setup utilizes a Varian 450-GC Gas Chromatograph with Pulsed Flame Photometric Detection (PFPD) to overcome matrix quenching issues found in single-channel systems.

What's inside

  • Introduction
  • Instrumentation
  • Technique
  • Materials and Reagents
  • Conditions
  • Results and Discussion
  • Repeatability data

Frequently asked questions

What sulfur components can be analyzed using this two-channel system?

The system is designed to analyze H2S, COS, methylmercaptane, ethylmercaptane, propylmercaptane, iso-propylmercaptane, tert-butylmercaptane, and tetrahydrothiophene.

What are the dimensions of the CP-Sil 5CB column used in the front channel?

The CP-Sil 5CB column measures 30 m x 0.32 mm with a film thickness of 5 µm.

What are the injection and oven temperature conditions for the analysis?

Injection is performed at 220 °C with a split ratio of either 1:30 or 1:20, while the oven ramps from 35°C to 250 °C.