Agilent SI-01663 Application Note Removal of Salts prior to Carbohydrate Analysis

This application note describes how to remove salts from samples using StratoSpheres™ PL-MIXED MP cartridges to protect HPLC columns.


Agilent SI-01663 Application Note Removal of Salts prior to Carbohydrate Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
115 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
C8BEC03776F9A6FB091DF3B5554AB0C7

About this manual

The document explains that enzymatic hydrolysis of celluloses and hemicelluloses requires buffer solutions, which must be desalinated before High-Performance Liquid Chromatography (HPLC) analysis to protect the column. The note details a method using StratoSpheres PL-MIXED MP SPE cartridges containing strong cation and anion exchange functionality.

What's inside

  • Introduction
  • Conditions
  • Figure 1: Phosphorus Concentration
  • Figure 2: Carbohydrate Concentration
  • Table 1: Recovery Results

Specifications

Cartridge Part NumberPL3540-C610
Bed Mass500 mg / 6.0 mL
Capacity1.2 mmol (0.6 mmol cations and 0.6 mmol anions)

Frequently asked questions

What is the purpose of removing salts before HPLC analysis?

Samples must be desalinated to protect the column from loaded cations like Pb2+, Ca2+, Na+, and K+.

Which cartridge is recommended for removing buffer ingredients?

The StratoSpheres PL-MIXED MP SPE strong anion and cation exchanger.

What is the recovery rate for glucose?

The recovery rate for glucose is 93%.