Varian Application Note SI-01235 Universal Detection of Combinatorial Libraries by HPLC and Low Temperature ELSD

This application note describes a method for detecting combinatorial libraries using high performance liquid chromatography (HPLC) and low temperature ELSD. The method can detect low molecular weight degradation products in DMSO without interference from DMSO. This method is also discussed for its use in highlighting unstable compounds in the early stages of drug discovery.


Varian Application Note SI-01235 Universal Detection of Combinatorial Libraries by HPLC and Low Temperature ELSD - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
204 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
172421561B168474A4F19F3813217C00

About this manual

The document explains that the Varian evaporative light scattering detector is universal and not dependent on optical properties, making it suitable for detecting compounds with weak or no UV chromophores. It highlights the utility of this method for identifying low molecular weight degradation products in DMSO without interference from the solvent itself.

The note details a specific application where organic compounds stored in frozen DMSO at 4 °C are analyzed to detect decomposition. By operating the Varian ELSD at 25 °C, it successfully detects impurities that would otherwise be masked by the DMSO peak or undetected by UV detection and mass spectrometry.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • Conditions
  • Results and Discussion
  • Conclusion

Frequently asked questions

What is the recommended flow rate for the HPLC analysis described in this note?

The recommended flow rate for the analysis is 1.0 mL/min.

At what temperature should the Varian ELSD be operated to detect low molecular weight degradation products?

The Varian ELSD should be operated at 25 °C to detect low molecular weight degradation products.

What are the settings for the nebulizer and evaporator temperatures in the described detection method?

The nebulizer temperature is set to 25 °C and the evaporator temperature is set to 25 °C.