Agilent ZORBAX 300SB-C8 Column Aging Application Note

This document introduces the effect of column aging with acidic mobile phase on ZORBAX 300SB-C8 (4.6 x 150 mm) (Agilent P/N: 883995-906) column.


Agilent ZORBAX 300SB-C8 Column Aging Application Note - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
257 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
7D7CF7DCF3CB86A6FAA52ED30E8DC718

About this manual

This document presents an application study titled 'Effect of Column Aging with Acidic Mobile Phase: Silanol-Sensitive Peptide Separation, pH 7'. It evaluates the ZORBAX 300SB-C8 (4.6 x 150 mm) column using a specific mobile phase gradient from 10-100% B over 45 minutes. The conditions include 10mM NH4H2PO4 at pH 7.0 for solvent A and 50 mM sodium perchlorate in 50% ACN / 50% H2O for solvent B, with a flow rate of 1 mL/min at 40°C.

What's inside

  • Effect of Column Aging
  • Acidic Mobile Phase
  • Silanol-Sensitive Peptide Separation
  • ZORBAX 300SB-C8
  • Charged Peptides
  • Silanol Accessibility
  • Application Study

Specifications

Column Dimensions4.6 x 150 mm
Part Number883995-906
Flow Rate1 mL/min
Temperature40°C
Detection Wavelength214 nm
Gradient Duration45 min

Frequently asked questions

Does silanol content change after exposure to low-pH eluent?

No, the silanol content does not change even after 18 liters of low-pH eluent.

What is the purpose of this test regarding peptides?

It serves as a very sensitive neutral-pH test of silanol accessibility.