Agilent ZORBAX 300SB-C8 (3.5 m) Micro Flow Cell Comparison Application Note

This document compares the use of micro and regular flow cells with ZORBAX 300SB-C8 (3.5 µm) column for peptide and protein analysis. It highlights the reduction of analysis time without compromising efficiency. By reducing the flow cell size, the effects of extra-column volume are minimized with small particle-size packing.


Agilent ZORBAX 300SB-C8 (3.5 m) Micro Flow Cell Comparison Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
296 KB
Published
03 July, 2012
Updated
28 September, 2023
MD5 checksum
3CA454D9A3868570702368E8A0EBC0D9

About this manual

This document presents a technical comparison between micro flow cells and regular flow cells when used with small particle-size column packings, specifically the ZORBAX 300SB-C8 (3.5 m). The analysis illustrates that using a micro flow cell reduces band broadening and improves peak resolution for proteins and peptides compared to a regular flow cell setup.

What's inside

  • Application
  • Micro Flow Cell
  • Regular Flow Cell
  • Comparison
  • Highlights
  • ZORBAX 300SB-C8
  • Proteins and Peptides
  • Band Broadening
  • Peak Resolution

Specifications

Column PackingZORBAX 300SB-C8 (3.5m)
Column Dimensions4.6 x 50 mm
Mobile Phase A95:5 Water:ACN with 0.10% TFA
Mobile Phase B5:95 Water:ACN with 0.085% TFA
Gradient10-60% B in 10 minutes
Injection Volume10pL
Flow Rate1 mL/min
DetectionUV( 215 nm)

Frequently asked questions

When is a micro flow cell suggested?

A micro flow cell is suggested when using small particle-size column packings.

What are the benefits of using a micro flow cell with ZORBAX 300SB-C8?

It reduces band broadening and improves peak resolution for proteins and peptides.

How does a micro flow cell affect analysis time?

Using a 50-mm column length with ZORBAX 300SB-C8 (3.5 m) can reduce the time of analysis for peptides and proteins without sacrificing efficiency.

How does reducing flow cell size affect column performance?

Reducing the flow cell size minimizes extra-column volume effects with small particle-size packing.