Agilent Oligonucleotides Fraction Collection Application Note

This document describes the collection of fractions from an oligonucleotide separation using gel electrophoresis and offline MALDI-TOF analysis. By using the Agilent oligonucleotide kit and Agilent Capillary Electrophoresis system, antisense oligonucleotides can be successfully separated. The Agilent ChemStation software allows for fully automated collection of fractions using signal A as the peak sensor. This method enables further analysis of the separated peaks and determination of the molecular mass.


Agilent Oligonucleotides Fraction Collection Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
65 KB
Published
03 July, 2012
Updated
28 September, 2023
MD5 checksum
01B9357716804C156EC42804D83B48F7

About this manual

This document details an application for separating antisense oligonucleotides using a sieving-polymer-filled capillary and collecting specific fractions for further analysis. The process utilizes the Agilent Capillary Electrophoresis system with ChemStation software to automate fraction collection based on peak detection.

Following separation, the collected fractions are subjected to offline MALDI-TOF mass spectrometry analysis. The text provides specific experimental conditions, including buffer types, voltage settings, and detection parameters, demonstrating the successful determination of analyte mass.

What's inside

  • Abstract
  • Experimental
  • Figure 1 Automated fraction collection
  • Figure 3 MALDI-TOF analysis
  • Conclusions
  • Equipment

Specifications

Capillary length24.5 cm
Capillary inner diameter100 µm
Injection time7s
Run voltage-25 kV
Detection wavelength260 nm/8 nm
Laser energy4 µJ
Polaritynegative

Frequently asked questions

What software enables automated fraction collection?

Agilent ChemStation software enables the fully automated collection of fractions using signal A as the peak sensor.

What is the determined mass of the collected fraction?

The determined mass of the collected fraction was 3230.10 Da.