Agilent Fast Ultra-Fast Analysis Using the Agilent 1200 Series Rapid Resolution LC System Compared to a Conventional Agilent 1100 Series LC System

Fast and Ultra-fast Analysis with the Agilent 1200 Series Rapid Resolution LC System Compared to a Conventional Agilent 1100 Series LC System Using Sub 2-µm Particle Columns


Agilent Fast Ultra-Fast Analysis Using the Agilent 1200 Series Rapid Resolution LC System Compared to a Conventional Agilent 1100 Series LC System - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
117 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
8EEA077857296179933606D32A3879B6

About this manual

This document presents an application note comparing the performance of the Agilent 1200 Series Rapid Resolution LC (RRLC) system against a conventional Agilent 1100 Series LC system. The comparison focuses on the ability to develop faster analytical methods using narrow-bore columns with sub-2-micron particles, which are critical for high-throughput laboratories and LC/MS compatibility.

The study highlights that the Agilent 1200 Series RRLC system offers superior performance due to its wider pressure and temperature ranges, lower delay volumes, and improved noise levels in the diode-array detector. Specifically, the 1200 Series can operate at higher flow rates and temperatures without exceeding pressure limits, resulting in shorter run times and cycle times compared to the 1100 Series.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and discussion
  • Experiments using a 4.6-mm ID column
  • Optimization of parameters
  • Performance comparison
  • Table 1

Specifications

Column internal diameter2.1 to 4.6 mm
Column lengthup to 150-mm
Column particle size1.8-pm
Maximum pressure600 bar
Temperature range10 degrees below ambient up to 100 °C
Detector data rate80 Hz

Frequently asked questions

What is the main advantage of the Agilent 1200 Series RRLC system over the 1100 Series?

The 1200 Series offers wider pressure and temperature ranges, lower delay volumes, and improved noise, allowing for faster methods and higher sample throughput.

Can the Agilent 1200 Series operate at 600 bar?

Yes, the Agilent 1200 Series RRLC system can be operated with up to 600 bar and up to 100 °C.

What column type was used in the comparison?

The comparison used ZORBAX SB C-18 columns packed with 1.8-pm particles.