Agilent 6330 Ion Trap LC/MS Application Note

Simultaneous assessment of drug metabolic stability and identification of metabolites using HPLC-Chip/ion trap mass spectrometry can effectively identify metabolites, save time and improve work efficiency


Agilent 6330 Ion Trap LC/MS Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
582 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
C57083C8C57AF015C947557F941C29FF

About this manual

This document describes a method for simultaneously measuring drug metabolic stability and identifying metabolites using automated nanospray LC/MS. The study utilizes the Agilent 6330 Ion Trap LC/MS coupled with an HPLC-Chip to analyze the drug buspirone at a concentration of 1 pM following incubation with rat liver S9 fractions. This approach allows for both assays to be performed in a single run, offering high sensitivity and efficiency compared to traditional methods that require separate incubations. The high sensitivity of the HPLC-Chip/MS approach enabled the detection of metabolites at very low substrate concentrations, which is critical for early-stage drug discovery.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • HPLC-Chip/MS system
  • Chromatographic conditions
  • Mass spectrometric conditions
  • Data analysis
  • Measurement of metabolic stability

Specifications

Flow rate0.3 pL/min
Injection volume1 pL
Drying gas flow3 L/min
Drying gas temperature300 °C
Capillary voltage-1800 V
Scan range100-500 m/z

Frequently asked questions

Which mass spectrometer model is used in this application?

The Agilent 6330 Ion Trap LC/MS is used for the analysis.

What substrate concentration is used for the metabolic stability assessment of buspirone?

The assessment is performed at a substrate concentration of 1 pM.

What is the incubation temperature and duration for the metabolic stability assay?

The incubation is carried out at 37 °C for 0 minutes and 28 minutes.

What is the flow rate used during the analytical pump gradient?

The flow rate was set to 0.3 pL/min.

Download PDF (582 KB)