Agilent 7500c ICP-MS Application Note for Toxic Elements in TCM

Determination of Toxic Elements in Traditional Chinese Medicine Using Inductively Coupled Plasma Mass Spectrometry


Agilent 7500c ICP-MS Application Note for Toxic Elements in TCM - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
92 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
7500c
MD5 checksum
F047700A0DBF7D0B9DFA72E0C20C370E

About this manual

This document details a method for analyzing 13 toxic elements (Be, Cr, Mn, Ni, Cu, Zn, As, Ag, Cd, Ba, Hg, Tl, and Pb) in Traditional Chinese Medicines (TCMs). The analysis utilizes Inductively Coupled Plasma Mass Spectrometry (ICP-MS) following microwave digestion for sample preparation. The method was validated using certified reference materials (shrub leaves and tea leaves) and standard addition recovery experiments, showing detection limits between 0.1 and 7.2 ng/g and recovery percentages between 90.3% and 109.7%.

What's inside

  • Introduction
  • Instrumentation
  • Reagents
  • Sample pretreatment
  • ICP-MS parameters
  • Calibration curves
  • Method detection limit
  • Determination of standard materials
  • Recovery
  • Sample analysis

Specifications

Instrument ModelAgilent 7500c ICP-MS
Microwave SystemCEM MARS5
Plasma Power1350 W
Plasma Gas Flow15.0 L/min
Auxiliary Gas Flow1.0 L/min
Target ElementsBe, Cr, Mn, Ni, Cu, Zn, As, Ag, Cd, Ba, Hg, Tl, Pb
Detection Limits Range0.1-7.2 ng/g

Frequently asked questions

Which elements are analyzed in this method?

The method analyzes 13 toxic elements: Be, Cr, Mn, Ni, Cu, Zn, As, Ag, Cd, Ba, Hg, Tl, and Pb.

What is the detection limit range for the target elements?

The method detection limits for all target elements were between 0.1 and 7.2 ng/g.

What recovery percentages were achieved in the validation?

The recovery percentages for all elements were between 90.3% and 109.7%.

Which certified reference materials were used for validation?

Two certified reference materials, shrub leaves (GBW07602) and tea leaves (GBW08518), were used.