Agilent 6210 TOF MS Application Note: Noncovalently Associated Complexes

This Application Note describes a method for studying noncovalent protein interactions using the HPLC-Chip/MS system coupled with an ESI-TOF mass spectrometer. The application note studies apomyoglobin and myoglobin protein complexes. The results show that ESI/MS is an efficient and accurate method for studying noncovalent protein interactions.


Agilent 6210 TOF MS Application Note: Noncovalently Associated Complexes - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
574 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
27C8A2DEE20ED2ADD8C748AF8CBC1087

About this manual

This document details a study analyzing noncovalently associated protein complexes, specifically apomyoglobin and myoglobin, using electrospray ionization mass spectrometry (ESI/MS). The study highlights the sensitivity, speed, and selectivity of the ESI/MS approach compared to traditional methods like ultracentrifugation and calorimetry. It demonstrates how the technique can identify protein complexes with high accuracy using small sample amounts.

What's inside

  • Introduction
  • Sample preparation
  • Analysis
  • TOF MS conditions
  • Results and Discussion
  • Apomyoglobin analysis
  • Heme b identification
  • Intact myoglobin characterization

Specifications

Mass accuracy< 2 ppm
Resolution>12,000
ESI voltage1650 V
Drying gas flow4.0 L/min
Drying gas temperature350°C
Fragmentor voltage275-300 V
Mass range400-3200 Daltons
Scan speed1 scan/second
Infusion flow rate600 nL/min

Frequently asked questions

What instrument was used to analyze the myoglobin complex?

An Agilent 6210 Time-of-Flight mass spectrometer equipped with an HPLC-Chip MS interface and Infusion Chip was used.

What is the mass accuracy of the TOF mass analyzer?

The TOF mass analyzer has a mass accuracy of less than 2 ppm.

How long does the analysis take?

The analysis is completed in less than five minutes.

What was the flow rate used for sample infusion?

Each sample was infused into the TOF at 600 nL/min using a syringe pump.