Agilent 1100 Series LC/MS Opiate Determination Application Note

This application note presents the procedure for the analysis of morphine and its metabolites in blood plasma using electrospray LC/MS, a method which does not require derivatization.


Agilent 1100 Series LC/MS Opiate Determination Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
11
File format
PDF
File size
160 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
504F9E711C1C7CA073DD798FCD86FA93

About this manual

This document details an analytical method for the determination of opiates and their metabolites in blood samples using electrospray LC/MS. The method utilizes the Agilent 1100 Series LC/MS system, which includes a binary pump, vacuum degasser, autosampler, thermostatted column compartment, diode-array detector, and the LC/MSD VL quadrupole mass spectrometer. The analysis is performed using electrospray ionization (ESI) in positive mode without the need for derivatization.

The procedure involves solid-phase extraction (SPE) for sample preparation, followed by chromatographic separation using a Supelco Discovery HSC18 column and a gradient mobile phase of water with 0.1% formic acid and methanol. The method achieves excellent recovery rates and accuracy, with coefficients of variation typically 5% or less. It allows for the identification and quantitation of various opiates, including morphine, hydromorphone, codeine, hydrocodone, 6-acetylmorphine, and oxycodone, as well as the internal standard nalorphine.

What's inside

  • Introduction
  • Experimental
  • Analytical standards
  • Sample preparation
  • Chromatographic Conditions
  • MS Conditions
  • Results and Discussion
  • Method accuracy and precision

Specifications

Mobile phase A0.1% formic acid in water
Mobile phase Bmethanol
Flow rate0.5 ml/min
Column temp50°C
Veap3000 V
Nebulizer40 psig
Drying gas flow13 l/min
Drying gas temp350°C
Mass rangem/z 100-650
Fragmentor120 V

Frequently asked questions

What system is used for the opiate analysis?

The Agilent 1100 Series LC/MS system, including a binary pump, vacuum degasser, autosampler, thermostatted column compartment, diode-array detector, and the LC/MSD VL quadrupole mass spectrometer.

What is the maximum flow rate used during the chromatographic analysis?

The flow rate for the analysis is set to 0.5 ml/min.

What temperature is used to dry the eluate after the sample cleanup step?

The eluate is evaporated to dryness with a stream of nitrogen at 40°C.