Agilent LC/MSD Trap Application Note: High-Throughput Quantitation of Midazolam in Human Serum

This document discusses the importance of target analysis of pharmaceutical compounds in biological matrices, and demonstrates the use of an ion trap LC/MS/MS system for fast and sensitive quantitation of the benzodiazepine drug midazolam in human serum.


Agilent LC/MSD Trap Application Note: High-Throughput Quantitation of Midazolam in Human Serum - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
5
File format
PDF
File size
90 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
3E762D6186A55F700C868A8B179AC3FD

About this manual

This document describes a method for the fast and sensitive analysis of the pharmaceutical compound midazolam in human serum. It highlights the advantages of modern ion trap mass spectrometers over traditional triple-quadrupole systems, specifically regarding speed, selectivity, and robustness in complex biological matrices.

The note details an experimental setup using an Agilent 1100 Series LC/MSD Trap with atmospheric pressure chemical ionization (APCI). Key results include a cycle time of only 2.0 minutes, a limit of quantitation of 32 ng/ml, and high linearity across a wide dynamic range.

What's inside

  • Introduction
  • Experimental Setup
  • Sample Preparation
  • Results and Discussion
  • Analysis Method Parameters
  • Conclusions

Specifications

ColumnZORBAX Eclipse XDB-C8, 50x2.1 mm, 3.5 µm
Flow rate500 µl/min
Mobile phase A0.1% acetic acid in water
Mobile phase BMethanol
Gradient modeIsocratic at 55% B
Ionization modePositive APCI
Nebulizer pressure35 psig
Corona current3500 nA
Limit of quantitation (Midazolam)32 ng/ml

Frequently asked questions

What is the limit of quantitation for midazolam?

The limit of quantitation for midazolam in serum is 32 ng/ml.

What is the overall cycle time for the analysis?

The overall cycle time is kept at only 2.0 minutes.

What ionization mode was used for this analysis?

The system was operated with an atmospheric pressure chemical ionization (APCI) source in the positive-ion mode.