Comparison of Sensitivity Spectra between the AP-MALDI LC/MSD Trap SL System and Traditional Vacuum MALDI-TOF Applicatio

Introduction to MALDI and its applications in protein and peptide analysis. Including the principle of MALDI, the analysis methods of MALDI, the advantages and disadvantages of MALDI


Comparison of Sensitivity Spectra between the AP-MALDI LC/MSD Trap SL System and Traditional Vacuum MALDI-TOF Applicatio - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
13
File format
PDF
File size
433 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
FFA81396B3F5BC0429C4ED2D57BC568E

About this manual

This document presents a technical comparison between the Agilent AP-MALDI ion source equipped on an LC/MSD Trap SL system and two commercial vacuum MALDI-TOF instruments. The study focuses on analyzing tryptic digests of proteins such as cytochrome c, lysozyme, myoglobin, and bovine serum albumin (BSA) at the femtomole level.

The results indicate that the Agilent AP-MALDI Trap system demonstrated superior signal-to-noise ratios compared to the vacuum MALDI-TOF systems. Additionally, the document highlights the advantage of generating true MS/MS data for peptide sequencing using the ion trap technology, which is not possible with standard reflectron TOF systems. The analysis also covers spectral similarities and differences, noting that AP-MALDI produces fewer low-mass ions due to softer ionization at atmospheric pressure.

What's inside

  • Introduction
  • Sample preparation
  • Instrument conditions
  • Sensitivity comparison
  • Spectral comparison
  • Peptide sequences identified
  • MS/MS capability
  • Conclusions

Frequently asked questions

What are the advantages of AP-MALDI-IT over vacuum MALDI-TOF?

AP-MALDI-IT decouples ionization factors from mass accuracy, permits true MS/MS data generation for peptide sequencing even at subfemtomole levels, and allows switching between modes in minutes.

What is the sensitivity level achieved by the Agilent AP-MALDI Trap?

The system enabled peptide analyses in the 250-500 attomole level, which is significantly below levels attained using many vacuum MALDI-TOF instruments.

Download PDF (433 KB)