Agilent 1100 Series LC/MSD Trap SL Positive Negative Mode Analyses Application Note

This document describes the polarity switching technology in mass spectrometric analysis and the ability of a mass spectrometer to switch polarities during a single chromatographic peak. This capability allows for the detection of various polarity molecules in unknown samples during a single run, saving time and improving data quality.


Agilent 1100 Series LC/MSD Trap SL Positive Negative Mode Analyses Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
5
File format
PDF
File size
73 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
29DA86236B827D339FD58A55736E895E

About this manual

This document describes the capabilities of the Agilent 1100 Series LC/MSD Trap SL ion trap mass spectrometer for analyzing samples containing compounds that favor both positive and negative ionization modes. It highlights the instrument's ability to switch polarities rapidly (approximately 0.5 seconds) to acquire data from unknown analytes in a single chromatographic run.

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Instrument Conditions
  • Conclusions

Specifications

Polarity switching time~ 0.5 seconds
Column temperature40°C
Flow rate0.4 ml/min
Injection volume3 pl
Capillary voltage3000 V
Scan range50-600 m/z

Frequently asked questions

What is the polarity switching speed of the Agilent 1100 Series LC/MSD Trap SL?

The instrument features extremely fast polarity switching of approximately 0.5 seconds.

What mobile phase additive was used for negative mode APPI in the example?

Acetone was added to the mobile phase to provide thermal electrons for negative mode APPI.