Agilent 1100 Series Real-time LC Run Mapping Application Note

Introduce a method for coupling nano-LC with MALDI mass spectrometry, which allows to deposit the LC run on the MALDI plate in real time, and thus couple the high resolution power of nano-RP-HPLC separation directly with MALDI mass spectrometry and decouple the aquisition of MS/MS from chromatographic separation and MS cycle time.


Agilent 1100 Series Real-time LC Run Mapping Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
699 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
BB6C592B0D2C2AF41C3905C3A1AE98A2

About this manual

This document details an optimized method for coupling nano-LC separation directly with MALDI mass spectrometry using the Agilent 1100 Series micro fraction collector. The system performs real-time LC run mapping on a MALDI plate, allowing for the deposition of separated fractions with online matrix addition. This approach decouples the acquisition of MS/MS analysis from the chromatographic separation cycle time.

The note outlines the experimental setup, including the use of an Agilent 1100 Series nanoflow pump, thermostatted autosampler, diode-array detector, and MSD Ion Trap XCT with AP-MALDI source. It explains the "liquid contact control mode" for reliable spotting and provides results from experiments optimizing solvent composition and plate temperature for protein identification.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Principle of spotting
  • MALDI sample preparation
  • Online matrix addition
  • Solvent composition optimization
  • Temperature optimization
  • Results

Specifications

Flow rate800 nL/min
Flow cell volume80-nL
Matrix solution composition50 % 2-propanol/49 % water/1 % acetic acid
Optimal plate temperature20 °C

Frequently asked questions

What is the principle of the spotting method?

The spotter ensures the capillary tip remains in contact with the surface of the deposited liquid on the MALDI target plate during the entire spotting process.

Which matrix solution resulted in the best protein identification?

A mixture of 2 mg/mL CHCA dissolved in 50 % 2-propanol/49 % water/1 % acetic acid resulted in the best protein identification.