Using ZORBAX Eclipse XDB-C18 Rapid-Resolution (3.5µm) Cartridge-Column, can simultaneously reduce column length and particle size, thereby reducing analysis time, maintaining separation, and reducing solvent usage.
This document describes an application for the rapid analysis of a pesticide mixture. It utilizes columns with particle sizes under 5 µm and dimensions up to ten times smaller than standard analytical columns to resolve six specific pesticides in less than two minutes.
The procedure uses an Agilent 1100 HPLC system equipped with a quaternary pump, a ZORBAX Eclipse XDB-C18 Rapid-Resolution (3.5µm) Cartridge-Column (4.6 x 30 mm), and UV detection at 254 nm. The method achieves high resolution while reducing analysis time and solvent consumption.
These columns are ideal for high-speed analyses that resolve six pesticides in less than two minutes.
No modification to a modern instrument plumbing or flow cell is necessary for this example.
The system uses an Agilent 1100 with a ZORBAX Eclipse XDB-C18 Rapid-Resolution (3.5m) Cartridge-Column (4.6 x 30 mm), MeOH: water (60:40) mobile phase, UV 254 nm detection with a 13 uL flow cell, 2mL/ min flow, and ambient temperature.