Agilent PLgel 10 um MIXED-B Columns Application Note for Polyvinyl Pyrrolidone Analysis with DMF

This document describes the application of Agilent PLgel 10 µm MIXED-B columns in high molecular weight polymer analysis and demanding eluent conditions, such as polyvinyl pyrrolidone in dimethylformamide (DMF). Polyvinyl pyrrolidone is soluble in both polar organics and water. DMF is used with LiBr to minimize any polyelectrolyte effects. The document also highlights the features of PLgel MIXED-B columns, which are particularly useful for molecular weight distributions where slightly higher than average MWs are encountered.


Agilent PLgel 10 um MIXED-B Columns Application Note for Polyvinyl Pyrrolidone Analysis with DMF - cover page
Brand
Analysis
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
529 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
84F2C9CFB27B63AD0214B5295F12B556

About this manual

Agilent Technologies presents an application note on analyzing Polyvinyl Pyrrolidone (PVP) dissolved in Dimethylformamide (DMF). The document highlights the suitability of Agilent PLgel 10 um MIXED-B columns for high molecular weight polymer analysis under demanding eluent conditions. It notes that PVP is soluble in polar organics and water, and DMF with Lithium Bromide (LiBr) is used to minimize polyelectrolyte effects.

What's inside

  • Introduction
  • Conditions

Specifications

Column Particle Size10 um
Molecular Weight Rangeup to 10 million
Flow Rate1.0 mL/min
Temperature60°C
Eluent CompositionDMF + 0.1% LiBr

Frequently asked questions

What polymer and solvent combination are these columns designed for?

The columns are designed for analyzing polyvinyl pyrrolidone using dimethylformamide (DMF) as the eluent.

What is the maximum molecular weight range covered by these columns?

The PLgel MIXED-B columns span a wide range of molecular weights up to 10 million.

What are the specific dimensions and particle size for the recommended column setup?

The recommended setup uses two columns with a length of 300 mm, an internal diameter of 7.5 mm, and a particle size of 10 µm.