Agilent VGA-76 Vapor Generation AA Application Note

This document introduces how to use vapor generation atomic absorption method to analyze arsenic, antimony and selenium in plant materials.


Agilent VGA-76 Vapor Generation AA Application Note - cover page
Brand
Analysis
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
165 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
VGA-76
MD5 checksum
542B8418EC71EAB56E34F8201570741E

About this manual

The document details a method for analyzing toxic elements like arsenic and antimony in plant materials using vapor generation atomic absorption spectroscopy. It highlights the use of a digestion procedure that avoids perchloric acid, utilizing a mixture of sulphuric, nitric, and hydrogen peroxide acids instead.

The study reports results obtained using an Agilent AA-1475 instrument equipped with background correction and a VGA-76 vapor generator. The analysis was performed on NBS orchard leaves (No. 1571), demonstrating the technique's ability to achieve results in good agreement with certified values.

What's inside

  • Introduction
  • Instrumentation
  • Sample Treatment
  • Results and Discussion
  • Summary
  • References

Specifications

Instrument ModelAgilent AA-1475
Vapor Generator ModelVGA-76
Recorder ModelAgilent 9176
Printer ModelHP 82905A
Sample TypeNBS orchard leaves (No 1571)
Sample Massapproximately 0.5 g
Digestion Volume5 mL concentrated sulphuric
Hydrogen Peroxide Volume5 mL volume of 30% v/v hydrogen peroxide
Nitric Acid Volume2 mL of concentrated nitric acid
Final Solution Volume25 mL
Final Acid Concentration7 M hydrochloric acid
Integration Timefive 3-second integrations

Frequently asked questions

Which acids are used in the digestion procedure to avoid perchloric acid?

The digestion procedure uses a mixture of sulphuric, nitric, and hydrogen peroxide acids.

What instrument was used to measure arsenic and antimony in the study?

An Agilent AA-1475 with background correction and a VGA-76 vapor generator were used.

How was the final solution prepared for analysis?

The digested mixture was made up to 25 mL to give a final solution in 7 M hydrochloric acid.

What calibration technique was necessary for the analysis?

The standard additions calibration procedure was necessary for the determination.

Download PDF (165 KB)