Agilent 1200/6200 Series Application Note: Structure Elucidation of Ginsenosides

This document describes the method of structure elucidation of complex natural product extracts from ginseng, using rapid resolution liquid chromatography and orthogonal time-of-flight mass spectrometry for component separation and accurate mass measurement. The document discusses the use of a rapid resolution liquid chromatography system and an ion trap mass spectrometer for structure elucidation.


Agilent 1200/6200 Series Application Note: Structure Elucidation of Ginsenosides - cover page
Brand
Analysis
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
194 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
3DB62D7431DA3AA004B819E94B95E3AD

About this manual

This document describes an analytical method for the structure elucidation of ginsenosides found in a complex Asian ginseng root extract. The study utilizes the Agilent 1200 Series Rapid Resolution LC system equipped with RRHT columns to achieve high-resolution separation of the extract's ingredients. Accurate mass measurements are performed using an Agilent 6200 Series orthogonal acceleration time-of-flight mass spectrometer (oaTOF) with an ESI interface.

The application details the specific experimental conditions for both the liquid chromatography and mass spectrometry systems, including solvent gradients, flow rates, and instrument parameters. Results show that the rapid resolution LC system provides superior peak resolution compared to standard systems, facilitating the identification of major constituents such as ginsenosides Rb1, Re, and Rg1 through accurate mass calculation and collision-induced dissociation (CID) experiments.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Sample Preparation
  • Methods
  • Results and Discussion
  • Experimental Conditions

Specifications

Binary Pump Max Pressure600 bar
Autosampler Delay Volume1.4 pL
Column Particle Size1.8 pm
DAD Sampling Rate80 Hz
MS Data Acquisition Rate40 Hz
Flow Rate0.5 mL/min
Column Temperature50° C
Injection Volume10 pL
Fragmentor Voltage150 V or 300 V

Frequently asked questions

What is the maximum pressure capability of the binary pump used?

The binary pump SL has the capability to deliver a pressure of up to 600 bar.

What is the particle size of the RRHT column used for separation?

The RRHT column used has a particle size of 1.8 pm.

Download PDF (194 KB)