This document details a High-Performance Liquid Chromatography (HPLC) method for analyzing warfarin, an anticoagulant drug. The method utilizes a Zorbax SB-C18 column and variable wavelength or fluorescence detection to determine the drug's concentration with high precision.

What's inside

  • Abstract
  • Chromatogram
  • Conditions
  • Performance
  • Equipment
  • Columns

Specifications

Column4.6 x 75 mm Zorbax SB-C18, 3.5 um
Mobile phase A0.025 M KH2PO4 in water (pH = 3 with H2SO4)
Mobile phase Bacetonitrile
Flow rate1.0 ml/min
Gradient20 % B to 80 % B in 10 min
Column compartment temperature25 °C
Injection volume5 ul
LOD0.1 mg/l

Frequently asked questions

What equipment is recommended for this analysis?

The method was specifically developed on the Agilent 1100 Series system.

What column is recommended for this method?

The recommended column is Zorbax SB-C18, 3.5 um, 4.6 x 75 mm.

What is the limit of detection for warfarin?

The limit of detection (LOD) for warfarin is 0.1 mg/l.