Agilent 1100 Series Analysis of Antiestrogen Drugs by HPLC Application Note

The HPLC method for analysis of antiestrogen drugs is described in this document.


Agilent 1100 Series Analysis of Antiestrogen Drugs by HPLC Application Note - cover page
Brand
Analysis
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
28 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
24E9D538B504294C07F3A33D9E7ED910

About this manual

This document presents an HPLC method for the analysis of antiestrogen drugs, specifically focusing on tamoxifen. The method utilizes a reversed-phase column and UV detection to inhibit the biosynthesis of estrogen, which is relevant for treating certain tumors like breast cancer. The analysis involves a gradient elution with a mobile phase of phosphate buffer and acetonitrile.

What's inside

  • Abstract
  • Conditions
  • Figure 1 Analysis of tamoxifen
  • Performance Table
  • Equipment
  • Columns

Specifications

ColumnPurospher RP-18 5 um 4 x 125 mm
Mobile phase A0.025 M KH2PO4 in water (pH = 3)
Mobile phase Bacetonitrile
Flow rate1.0 ml/min
Gradient20 % B to 60 % B in 14 min
Column wash60 % B to 20 % B in 2 min
UV detector wavelength204 nm
Column temperature60 °C
Injection volume5 ul
Stop time16 min
Post time5 min
LOD for Tamoxifen0.2 mg/l

Frequently asked questions

What is the purpose of the HPLC method described?

The method is used to analyze antiestrogen drugs, specifically tamoxifen, to inhibit the biosynthesis of estrogen from testosterone.

What is the recommended column for this analysis?

The recommended column is Purospher RP-18, 5 um, 4x 125 mm.

What is the flow rate for the HPLC analysis?

The flow rate is set to 1.0 ml/min.

What is the limit of detection (LOD) for tamoxifen?

The LOD for tamoxifen is 0.2 mg/l.