Varian Application Note SI-01213 Effective HPLC Determination of Fentanyl using ELSD

This application note describes a method for the effective HPLC determination of fentanyl using an ELSD detector.


Varian Application Note SI-01213 Effective HPLC Determination of Fentanyl using ELSD - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
274 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
FDF6C17083937CE0BA79FC66B3230CA4

About this manual

This document details an HPLC method for the determination of fentanyl, a potent analgesic and anaesthetic. The method utilizes a Varian evaporative light scattering (ELS) detector, which is noted for its superior performance and independence from the optical properties of the compound. The study analyzes fentanyl concentrations ranging from 0.5 to 200 µg/mL.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • Results and Discussion
  • Conclusion

Specifications

DetectorVarian ELSD
Nebulizer Temperature40 °C
Evaporator Temperature70 °C
Gas Flow1.2 SLM
Eluent Composition60 % 0.25 M Ammonium acetate in water, 40 % Acetonitrile
Flow Rate0.25 mL/min
Injection Volume10 µL
Limit of Detection10 ng on-column loading

Frequently asked questions

What eluent is used for the fentanyl analysis?

The eluent consists of 60% 0.25 M Ammonium acetate in water and 40% Acetonitrile.

What is the detection limit for fentanyl in this method?

The limit of detection is 10 ng on-column loading.

What are the flow rate and injection volume for the HPLC analysis?

The flow rate is 0.25 mL/min and the injection volume is 10 µL.