Varian 220-MS Ion Trap GC/MS/MS Application Note SI-02407

This application note describes a method using the Varian 220-MS ion trap mass spectrometer (Figure 2) to determine melamine and its analogs qualitatively and quantitatively based on the framework of the original FDA method.


Varian 220-MS Ion Trap GC/MS/MS Application Note SI-02407 - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
2.3 MB
Published
12 June, 2012
Updated
30 September, 2023
Model
220-MS
MD5 checksum
48D81459E3B58E95A00191A695B799B2

About this manual

This document details a method developed to determine melamine, ammeline, ammelide, and cyanuric acid in infant milk-based formula and other food/feed products. The method utilizes the Varian 220-MS ion trap mass spectrometer coupled with a gas chromatograph and V:Results software to perform qualitative and quantitative analysis.

What's inside

  • Introduction
  • Instrumentation
  • Methods and Materials
  • Extraction Procedure
  • Standard Curve
  • GC Conditions
  • MS/MS Parameters
  • Results and Discussion
  • Recovery studies
  • Conclusions

Specifications

ColumnFactorFour VF-Xms, 30 m x 0.25 mm x 0.25 pm
Inlet Temperature280 °C
Carrier Gas FlowHelium at 1 mL/min
Trap Temp220 °C
Detection Limitbelow 10 ppb

Frequently asked questions

What compounds can this method detect?

The method detects melamine, cyanuric acid, ammelide, and ammeline.