This document describes the method of fast fingerprint analysis of perfumes using Agilent J&W FactorFour VF-WAXms high efficiency GC columns. The column shows excellent selectivity for the volatiles present in perfumes and allows significant reduction in analysis time without compromising on resolution of separation.


Agilent Fast Fingerprint Analysis of Perfume using 0.15 mm VF-WAXms GC Columns - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
250 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
674D8A16CE14E054AC3A972B9D2A9C0D

About this manual

This document presents an application note by Helena Jacobse from Agilent Technologies regarding the use of VF-WAXms columns for analyzing perfumes. The text highlights that the 0.15 mm column allows for significantly reduced analysis time without compromising separation resolution compared to the 0.25 mm column. It includes a comparison of chromatograms showing the selectivity for volatiles present in perfumes.

What's inside

  • Introduction
  • Conditions
  • Figure 1

Specifications

Column TypeVF-WAXms High Efficiency GC Columns
Column Dimensions (Fast)15 m x 0.15 mm df = 0.15 mm
Column Dimensions (Standard)30 m x 0.25 mm df= 0.25 um
Carrier GasHelium, 2.0 bar
Temperature Range75 °C to 260 °C
Temperature Ramp16 °C/min
Injector Temperature250 °C
Detector Temperature275 °C
Sample Size1.0 ub
Sample CompositionPerfume, 2 % (acetone)

Frequently asked questions

What is the benefit of using the 0.15 mm VF-WAXms column?

It allows analysis time to be reduced significantly without compromising on the resolution of the separation.

What sample was used in the perfume analysis?

The sample used was perfume diluted with 2% acetone.

Download PDF (250 KB)

More Agilent Manuals


View all Agilent manuals →