Agilent 2236 Vitamin B6 Complex Separation Application Note

This document describes the separation method of Vitamin B6 complex on Pursuit Diphenyl column. The mobile phase consists of 5 mM ammonium formate and CH3CN in a ratio of 95:5. The flow rate is 0.5 mL/min and detection is performed at 254 nm. This method is suitable for the separation and detection of pyridoxamine, pyridoxal, and pyridoxine.


Agilent 2236 Vitamin B6 Complex Separation Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
1
File format
PDF
File size
27 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
7198FF2FFD5E9FCCD089087514DA4A24

About this manual

This document provides an application note for the Agilent 2236 Liquid Chromatography system. It details the specific method for separating a Vitamin B6 complex, which includes Pyridoxamine, Pyridoxal, and Pyridoxine.

The recommended method utilizes a Pursuit Diphenyl column with dimensions of 250 x 4.6mm and a 5-micron particle size. The mobile phase consists of 5mM Ammonium formate at pH 2.65 mixed with Acetonitrile (MeCN) in a 95:5 ratio. The analysis is performed at a flow rate of 0.5 mL/min and a temperature of 40°C, with detection at 254 nm.

What's inside

  • Vitamin B6 complex separation
  • Pursuit Diphenyl Column
  • Mobile Phase composition
  • Flow Rate and Temperature
  • Detection wavelength
  • Component identification

Specifications

ColumnPursuit Diphenyl, 250 x 4.6mm, 5u
Mobile Phase5mM Ammonium formate, pH 2.65: MeCN - 95:5
Flow Rate0.5 mL/min
Temperature40°C
Detection254 nm

Frequently asked questions

What components are separated in the Vitamin B6 complex?

The complex includes Pyridoxamine, Pyridoxal, and Pyridoxine (or Pyridoxol or Vitamin B6).

What is the recommended flow rate for this separation?

The recommended flow rate is 0.5 mL/min.