Agilent Polaris 5u C8-Ether Column Application Note for Analgesics Analysis

This application note describes a procedure for the separation of four common analgesics by high performance liquid chromatography (HPLC) using a Polaris 5µ C8-Ether column. A 0.1% formic acid solution was used as the mobile phase. The four analgesics can be clearly seen in the chromatogram.


Agilent Polaris 5u C8-Ether Column Application Note for Analgesics Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
1
File format
PDF
File size
42 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
A06B10C7CA9DB3714537472FE098934C

About this manual

This document provides a liquid chromatography method for identifying specific analgesic compounds. The procedure utilizes an Agilent Polaris 5u C8-Ether column with dimensions of 150 x 4.6mm (Part Nr.: A2030150X046).

The analysis employs a gradient elution starting at 40% and increasing to 90% of solvent B over 10 minutes, with a flow rate of 21.0 ml/min. Detection is performed using UV detection at 220nm.

What's inside

  • Analgesics
  • Polaris 5u C8-Ether Column
  • Gradient Elution
  • Peak Identification
  • Acetaminophen
  • Acetylsalicylic Acid
  • Salicylic Acid
  • Ibuprofen

Specifications

Column Part NumberA2030150X046
Column Dimensions150 x 4.6mm
Solvent A CompositionH2O + 0.1% HCOOH
Solvent B CompositionMeCN + 0.1% HCOOH
Gradient Program40% to 90%B in 10 min
Flow Rate21.0 ml/min
Detection WavelengthUV @ 220nm

Frequently asked questions

Which analgesics are identified in this application?

The document lists acetaminophen, acetylsalicylic acid, salicylic acid, and ibuprofen.

What is the total run time for this analysis?

The gradient elution runs for a total of 10 minutes.