Agilent Polaris 5u C18-A Antidiabetics Application Note

The document is about application note, it introduces the application method and the instruments and reagents used.


Agilent Polaris 5u C18-A Antidiabetics Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
1
File format
PDF
File size
29 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
EF79B2CCBF44327F5B55A086657DCF48

About this manual

This document is an application note from Agilent Technologies regarding the analysis of antidiabetics. It references the acquisition of Varian, Inc. by Agilent and directs users to the company website for further information.

The note specifies the use of the Polaris 5u C18-A column (Part nr. A2000150X046) with dimensions of 150 x 4.6mm. The recommended eluent is a mixture of MeCN and 75 mM (NH4)3PO4 at pH 4.4 in a 70:30 ratio, with a flow rate of 21.0 ml/min and UV detection at 254nm. The chromatogram identifies glyburide as a peak.

What's inside

  • Antidiabetics
  • Polaris 5u C18-A
  • Chromatographic Conditions
  • Peak Identification
  • Glyburide

Specifications

ColumnPolaris 5u C18-A, 150 x 4.6mm
Part NumberA2000150X046
EluentMeCN : 75 mM (NH4)3PO4, pH 4.4, 70 : 30
Flow Rate21.0 ml/min
DetectionUV @ 254nm

Frequently asked questions

Which column is used for the antidiabetics analysis?

The Polaris 5u C18-A column is used.

What is the part number for the specified column?

The part number is A2000150X046.

What is the flow rate for the analysis?

The flow rate is 21.0 ml/min.

Download PDF (29 KB)