About this manual

This document provides an HPLC method for the analysis of beta-blockers. The technique utilizes a Pursuit 5 C8 column (150 x 4.6mm) with a ChromSep Cartridge System. The analysis is performed at ambient temperature using a mobile phase consisting of MeCN and 0.1% NH4OH in water (pH 10.3) at a flow rate of 21.0 mL/min.

What's inside

  • Notice
  • Application 2326 LC
  • Peak Identification

Specifications

ColumnPursuit 5 C8, 150 x 4.6mm
Part no.Conventional A3030150X046
SystemChromSep Cartridge System A3030150C046
TemperatureAmbient
Eluent AMeCN
Eluent B0.1% NH4OH in water, pH 10.3
Ratio A:B65:35
Flow rate21.0 mL/min
DetectionUV @ 215 nm

Frequently asked questions

Which beta-blockers are identified in this application note?

The identified compounds are Hydrochlorothiazide, Practolol, Pindolol, Bisoprolol, and Alprenolol.

What is the flow rate for this analysis?

The specified flow rate is 21.0 mL/min.

At what wavelength does detection occur?

Detection takes place at a UV wavelength of 215 nm.