Hanna Instruments HI 2300 Manuals

ManualLib has 2 documents for the Hanna Instruments HI 2300 (Analyzer): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
HI 2300
Category
Analyzer
Documents
2
Languages
EN
Last added
2023-11-16

Start here

Hanna Instruments HI 2300 EC/TDS/NaCl Bench Meter Instruction Manual

Hanna Instruments HI 2300 EC/TDS/NaCl Bench Meter Instruction Manual

User Manual 36 pages · 1.6 MB · EN

The Hanna Instruments HI 2300 is a microprocessor-based bench meter that measures conductivity (EC), total dissolved solids (TDS), sodium chloride (NaCl), and temperature. It features automatic or manual temperature compensation, a user-selectable temperature coefficient, and an auto-ranging function for EC and TDS scales. The meter includes Good Laboratory Practice (GLP) data tracking and can store up to 500 logged readings in non-volatile memory.

The instrument is operated via a front-panel keypad with dedicated keys for calibration, range selection, logging, and setup. It connects to a PC through a USB interface for data transfer. The package includes a HI 76310 conductivity/TDS probe, a 12 VDC power adapter, and the instruction manual. The meter is warranted for two years, while probes carry a six-month warranty.

Read online

Hanna Instruments HI 2300 specifications

Measurement parametersConductivity (EC), TDS, NaCl, Temperature
Memory capacity500 manually logged points
PC interfaceUSB
Power supply12 VDC adapter
Temperature compensation modesAutomatic (ATC), Manual (MTC), No compensation (noTC)
Default temperature coefficient1.90 %/ºC
Auto-rangingEC and TDS scales auto-range
Included accessoriesHI 76310 probe, 12VDC adapter, instruction manual

Source: Hanna Instruments HI 2300 EC/TDS/NaCl Bench Meter Instruction Manual

All Hanna Instruments HI 2300 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

How many data points can the HI 2300 store?

The meter’s memory can hold 500 manually logged points.

From: Hanna Instruments HI 2300 EC/TDS/NaCl Bench Meter Instruction Manual

What temperature compensation modes are available?

Automatic (ATC), manual (MTC), or no compensation (noTC).

From: Hanna Instruments HI 2300 EC/TDS/NaCl Bench Meter Instruction Manual

What is the default temperature compensation mode?

The default compensation mode is ATC (Automatic Temperature Compensation).

From: Hanna Instruments HI 2300 Bench Meter Instruction Manual

How do I disable autoranging to freeze a specific range?

Press the LOCK key to disable autoranging and freeze the current range on the LCD.

From: Hanna Instruments HI 2300 Bench Meter Instruction Manual