Hanna Instruments LP2000-11 Precision Bench Turbidity Meter Instruction Manual
The Hanna Instruments LP2000-11 is a microprocessor-based bench top instrument designed to determine the turbidity of water and wastewater. It operates according to ISO 7027 standards using an infrared LED light source with a wavelength peaking at 890 nm, which minimizes interference from colored samples.
Read onlineHanna Instruments LP 2000 specifications
| Range | 0.00 to 50.00 FTU; 50 to 1000 FTU |
|---|---|
| Resolution | 0.01 / 1 FTU |
| Accuracy | 0.5 FTU + 5% of reading (whichever greater) |
| Light Source | High-emission infrared LED |
| Power Supply | 12 Vdc power adapter |
Source: Hanna Instruments LP2000-11 Precision Bench Turbidity Meter Instruction Manual
All Hanna Instruments LP 2000 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How many points can be used for calibration?
The meter supports two or three point calibration using standards at 0, 10, and 500 FIU.
From: Hanna Instruments LP2000-11 Precision Bench Turbidity Meter Instruction Manual
What is the maximum number of samples that can be stored in memory?
The instrument can store up to 500 samples before displaying a 'FULL' warning.
From: Hanna Instruments LP2000-11 Precision Bench Turbidity Meter Instruction Manual
What standard is used for calibration?
The instrument uses the AMCO-AEPA-1 primary standard to avoid formazine-related problems.
From: Hanna Instruments LP 2000 Precision Bench Turbidity Meter Instruction Manual
How often should calibration be performed?
Calibration is recommended at least once a month using the 10 FTU standard.
From: Hanna Instruments LP 2000 Precision Bench Turbidity Meter Instruction Manual