Key facts
- Brand
- Hanna Instruments
- Model
- HI-96721
- Category
- Other Electronics
- Documents
- 2
- Languages
- EN
- Last added
- 2012-07-10
Start here

Hanna Instruments HI 96721C Iron High Range ISM Instruction Manual
The HI 96721 is an auto-diagnostic portable microprocessor meter designed for measuring total iron (Fe) content in water and wastewater. It utilizes a special tungsten lamp and narrow band interference filter based on the USEPA Method 315B and Standard Method 3500-Fe B to ensure accurate and repeatable readings.
The instrument features an advanced optical system, splash waterproof design, and auto-shut off capabilities. It includes a CAL CHECK validation function using NIST traceable standards to verify performance at any time. The meter operates with a single 9V battery and measures iron concentrations in the range of 0.00 to 5.00 mg/L.
Hanna Instruments HI-96721 specifications
| Range | 0.00 to 5.00 mg/L |
|---|---|
| Accuracy | +0.04 mg/L + 2% of reading @ 25°C |
| Light Source | Tungsten lamp |
| Weight | 290 g (10 oz) |
Source: Hanna Instruments HI 96721C Iron High Range ISM Instruction Manual
All Hanna Instruments HI-96721 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What method does the HI 96721C use for measurement?
The meter uses an adaptation of the USEPA Method 315B and Standard Method 3500-Fe B.
From: Hanna Instruments HI 96721C Iron High Range ISM Instruction Manual
How long does the auto-shut off feature last?
The instrument turns off after 10 minutes in measurement mode or 1 hour in calibration mode.
From: Hanna Instruments HI 96721C Iron High Range ISM Instruction Manual
What is the measurement range of the HI-96721?
The meter measures total iron content in water samples in the 0.00 to 5.00 mg/L (ppm) range.
From: Hanna Instruments HI-96721 Iron High Range ISM Instruction Manual
What methods does this instrument adapt?
The method is an adaptation of USEPA Method 315B for natural and treated waters, and Standard Method 3500-Fe B for water and wastewater.