Hanna Instruments HI 96724 Manuals

ManualLib has 2 documents for the Hanna Instruments HI 96724 (Other Electronics): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
HI 96724
Documents
2
Languages
EN
Last added
2012-07-10

Start here

Hanna Instruments HI 96724C Free & Total Chlorine ISM Instruction Manual

Hanna Instruments HI 96724C Free & Total Chlorine ISM Instruction Manual

User Manual 24 pages · 1.4 MB · EN

The HI 96724 is an auto-diagnostic portable microprocessor meter designed for measuring free and total chlorine content in water. It features an advanced optical system with a tungsten lamp and narrow band interference filter, ensuring accurate and repeatable readings according to USEPA Method 330.5 and Standard Method 4500-Cl G.

The instrument includes user-friendly functions such as CAL CHECK validation, GLP mode for data integrity, and an auto-shut-off feature. It operates using a positive-locking cuvette system and requires specific reagents (HI 93701A-T, B-T, or C-T) to perform measurements in the range of 0.00 to 5.00 mg/L.

Read online

Hanna Instruments HI 96724 specifications

Range0.00 to 5.00 mg/L of Cl2 (Free or Total)
Resolution0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L
Accuracy+/-0.03 mg/L +3% of reading @ 25°C
Light SourceTungsten lamp
Environment0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type1x9 volt
Auto-Shut off (Measurement)After 10' of non-use in measurement mode

Source: Hanna Instruments HI 96724C Free & Total Chlorine ISM Instruction Manual

All Hanna Instruments HI 96724 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the method used for measuring chlorine?

The meter uses an adaptation of the USEPA Method 330.5 for wastewater and Standard Method 4500-Cl G for drinking water.

From: Hanna Instruments HI 96724C Free & Total Chlorine ISM Instruction Manual

What reagents are required for the test?

The instrument requires HI 93701A-T, HI 93701B-T, or HI 93701C-T reagents depending on the specific chlorine type being measured.

From: Hanna Instruments HI 96724C Free & Total Chlorine ISM Instruction Manual

What is the measurement range of the HI 96724?

The instrument measures free or total chlorine concentrations from 0.00 to 5.00 mg/l.

From: Hanna Instruments HI 96724 Free & Total Chlorine ISM Instruction Manual

How long should the instrument wait before taking a total chlorine reading?

After adding the sample, you must wait for 2 and 30 seconds before pressing the READ key.

From: Hanna Instruments HI 96724 Free & Total Chlorine ISM Instruction Manual