Key facts
- Brand
- Hanna Instruments
- Model
- HI 98121
- Category
- Analyzer
- Industry
- Industrial
- Documents
- 2
- Languages
- EN · ES
- Last added
- 2026-11-09
Start here

Hanna Instruments HI 98121 Water Ionizer pH/ORP Tester Supplemental Instruction Manual
This supplemental instructional manual from Hanna Instruments covers use of the HI 98121 portable pH and ORP tester for water ionizer applications. It answers frequent questions about why the tester may stop reading high negative mV ORP values and how often the pH/ORP sensor should be cleaned in HI 7061 cleaning solution.
The manual explains the pH scale from 0 to 14 and the meaning of ORP in millivolts, then describes ORP preparation and measurement, including pretreatment with HI 7091M or HI 7092M solutions and a pH/mV reference table. Maintenance, cleaning, polishing of the platinum sensor and storage with HI 70300M storage solution are also described, along with the solutions contained in the ionizer solutions kit.
Hanna Instruments HI 98121 specifications
| Product Type | Medidor Impermeable de pH / ORP y temperatura |
|---|---|
| Temperature Compensation | Automatic (ATC) |
| Calibration | One or two points with automatic buffer recognition |
| Included Electrode | HI 73127 |
| Included Tool | HI 73128 |
| Battery Type | 4 x 1.5V |
| Warranty Period | 1 year |
Source: Hanna Instruments HI 98121 pH/ORP and Temperature Meter User Manual
All Hanna Instruments HI 98121 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
Why doesn't my tester read high -mV ORP (-250mV to -999mV) anymore?
Mineral buildup on the surface of the platinum ORP sensor interferes with the full -mV scale. Try cleaning the sensor in HI7061 cleaning solution for 15 minutes.
From: Hanna Instruments HI 98121 Water Ionizer pH/ORP Tester Supplemental Instruction Manual
How often should I clean my pH/ORP sensor in HI7061 cleaning solution?
Clean the pH/ORP sensor after each use by soaking it in HI7061 cleaning solution for 2-3 minutes.
From: Hanna Instruments HI 98121 Water Ionizer pH/ORP Tester Supplemental Instruction Manual
What is the warranty period for the HI 98121?
The instrument is guaranteed for one year against manufacturing and material defects.
From: Hanna Instruments HI 98121 pH/ORP and Temperature Meter User Manual
What happens if the battery voltage is too low?
The meter displays '0%' and automatically disconnects via the Battery Error Prevention System (BEPS).
From: Hanna Instruments HI 98121 pH/ORP and Temperature Meter User Manual